Customised Gas chromatography

State-of-the-art, microprocessor controlled desktop GC’s, with 21 CFR-11 software.

OVEN

  • Dimensions: 27 x 25 x 30 (W x D x H) cms
  • Volume : 21 liters appx.
  • Temperature Range : 5°C above ambient to 450°C in steps of 1°C.
  • Ramp Rate : 0° to 50°C/min in 1°C/min increments.
  • Temperature Accuracy : ± 0.1°C
  • Safety interlock: Overheat protection, adjustable from 50°C to 450°C.
  • Built-in Alarm and Auto shut off.

PNEUMATICS
Column carrier gas regulation can be set within ±0.1 PSI.
Digital Column back pressure display.
Pressure Range : 0 – 60 PSI
Pressure Accuracy : ± 0.1 PSI
Digital  Control of :
» Capillary Column : Pressure
» Packed Column : Digital Flow Setting with Imported DFC
» Makeup Gas for FID : Digital Pressure Setting
» Hydrogen for FID : Digital Pressure Setting
» Oxidant / Air for FID : Digital Pressure Setting
» Flame Ionisation Detector : Electronic control of Range and Auto Zero through software.

INJECTORS: We can offer you twelve different injectors. The GC can accommodate upto three injection ports, packed or capillary, in any combination.

DETECTORS:  With a range of ten different detectors, we can meet any analytical need you might have.

SOFTWARE:  Choose from three software packages, depending on your data processing needs.